- DME CERTIFIED
- SUBJECT TO AVAILABILITY
- Software revision 6.51
- 40GB Hard drive
- Thermal printer (OLD STYLE)
- Lighted Work Station
- Custom Cart with High end wheels and drawer
- Paper tray and Acquisition module holder
- Custom Protocols
- PreProgrammed stress protocols
- Flat Screen
- Impedance Testing to verify patient hook up
- Automatic ecg strips and Blood Pressure Prompts
- Compatible with ge T2000 and T2100 treadmills
- Windows XP
Singal Processing:
- ECG Analysis Frequency: 500 Hz
- ST Measurements: ST amplitudes, slope, integral, index, ST/HR slope, ST/HR loops, ST/HR index up to 15 leads
- E, J and Post-J Point: Manual or computer selected
- Signal Processing: Incremental median updating technique
- Baseline Correction: Finite Residual Filter (FRF) and/or Cubic Spline analysis
- QRS Detection and Analysis: Automatic or manual lead selection
- ECG Output: Real-time ECG/QRS beep/TTL synchronization output
- Heart Rate: Automatic arrhythmia detection, documentation and annotation
- Full-Disclosure ECG: Beat-to-beat ECG storage & event review
- Reanalysis: Post-test medians measurements from E, J and post-J point selections
- ECG Interpretation (Optional): Marquette* 12SL* resting ECG analysis program for adults
and pediatrics
- Additional ECG Function: XTI stress interpretation; Vectorcardiography (15 lead)
Data Aquisition:
- Technology: Active, “Type BF” floating isolated powered 14-channel acquisition module with built-in lead-fail detection and lead prep impedance measurement
- Sampling Rate: Over-sampling @ 4000 Hz
- Dynamic Range: 320 mV, ± 10 mV signal superimposed on ± 150 mV DC offset
- Resolution: 4.88 μV/LSB @ 500 Hz
- Noise: < 15 μV peak-to-peak noise over 0.01 to 150 Hz (-3 dB) bandwidth
- Frequency Response: -3 dB, display and writer
- High-Pass Filter: 0.01 (or 0.05 Hz, special use) with DC offset control
- Low-Pass Filter: 20, 40, 100, 150 Hz (selectable)
- Line Filter: 50.0 or 60.0 Hz notch filter (selectable)
- Baseline Correction: Cubic Spline algorithm
- Artifact/Baseline Correction: Finite Residual Filter (FRF) analysis
- Common Mode Rejection: > 140 dB (123 dB with AC filter disabled)
- Input Impedance: > 10 M Ohms @ 10 Hz, defibrillator protected
- Patient Leakage: < 10 μA
- Pace Detect: Orthogonal LA, LL and V6; 750 μV @ 50 μs
Communications/Storage Options:
- MUSE* systems compatible via diskette; network (optional)
- MUSE Web compatible for retrieval view and printing of MUSE system data
- Local storage: Minimum 40 GB hard drive storage of complete ECG record and test results
- PDF export of final reports with auto naming protocols also from Microsoft® Word and Excel
- XML export of specified data
- CASE* network server with editing workstations
- Windows® NT
Display:
- Display Type: LCD (flat panel display)
- Display Resolution (Minimum): LCD-1280 x 1024
- Display Size: 17 or 19-inch LCD
- Monitored Leads: 3, 6, 12 or 15
- Displayed Leads: Number on screen 3, 6, 12 or 15
- Display Format: 3 rhythm, 3 rhythm + medians, 3 rhythm + trends, 6 rhythm, 4 x 2.5 + 1 rhythm, 2 x 6
- Display Speeds 25, 50 mm/s
- Display Sensitivity/Gain: 2.5, 5, 10, 20 mm/mV
- Displayed vital Signs Data (Configurable): Heart rate, target heart rate, blood pressure, exercise clock, stage clock, phase clock, protocol, speed, grade, Watts, METS, RRP and SpO2
- Displayed data ST scan/median complexes, arrhythmias, ventricular ectopic/min counter, 3 to 12 lead waveforms, lead check torso and 12 leads, waterfall display, trends, tabular summary, stored ECG strips, interpretation, time-of-day clock, patient name, warning messages and prompts
Writer:
- Writer Technology: “Instant” load, thermal dot array
- Writer Leads: 3, 6, 12 or 15 leads (standard, NEHB, Cabrera, configurable)
- Writer Speeds: 5, 12.5, 25 and 50 mm/sec (± 2%)
- Writer Sensitivity/Gain: 2.5, 5, 10 or 20 mm/mV (± 5%)
- Writer Resolution: Horizontal 1000 lines/sec x 200 dpi dedicated local printing. 200 x 200 dpi generic printing
- Paper Type: Thermal, perforated, fan-fold, 300 sheets/pack
- Paper Size: A size 214.63 x 280 mm (8.45 x 11 in); A4 size 210 x 297.5 mm (8.27 x 11.0 in)
Physical Specifications:
- Height: 148 cm (58 in) with no monitor installed
- Width: 62 cm (24 in) with paper tray removed
- Depth: 75 cm (30 in)
- Weight 68 kg (150 lb) without monitor
Minimum Server Requirements:
- Microprocessor: Pentium® ! 1.6 GHz
- RAM: 256 MB
- Hard Drive: > 40 GB
- Operating system: Windows NT
Interfaces Included:
- Acquisition module
- Keyboard (PS/2) anti dedicated stress
keypad (USB)
- Mouse (PS/2)
- Built-in thermal printer (USB)
- 100 Mbps ethernet MUSE compatible
- 6 serial ports: (COM 1-2, COM A-D)
treadmill, BP, ergometer, SpO2
- 4 analog and 1 TTL (trigger) output
analog ergometer, camera sync, etc.
- Diskette drive: 3 1/2 in, 1.44 MB
- PC-compatible for data storage
- CD-R/W drive: readable for software
updates; archivable data optional
Electrical Power:
- Power Supply: AC operation only
- Operating voltage range 100-120 VAC, 50-60 Hz, 2.8 A; 200-240 VAC, 50-60 Hz, 1.4 A
- Power consumption 350 W max (1200 BTU/hr); < 250 W normal (850 BTU/hr); < 30 W standby (100 BTU/hr)
Opperating Conditions:
- Temperature: +10 to 40°C (+50 to 104°F)
- Humidity: 20 to 95% RH non-condensing
Storage/Transport Conditions:
- Temperature -40 to +70°C (-40 to 158°F); Paper discoloration may occur at high temperatures
- Humidity: 15 to 95% RH non-condensing