Streamline your ECG workflow. With the MAC 5 you can complete a high-quality exam in just a few clicks and send the results anywhere, quickly and securely. All with minimal training or support.
Everything about the MAC 5 experience is designed to keep you moving forward and keep the focus on patient care. Quickly acquire a quality ECG. MAC 5 empowers every ECG professional to accelerate the care process.
We mean it when we say MAC 5 is Simple. Precise. Secure.
- Enhanced Hookup Advisor
- 8.9 inch touchscreen
- Built-in PDF/XML export support
- Bidirectional HL7 and DICOM communication
- NO PRINTER
- Optimal mobility
- Easy-clean design
- Auto-ECG3 algorithm
- Smart Lead technology
- Over 200 scientific references
- Gender- and pediatric-specific interpretations
- Full disclosure
- Advanced algorithms
- Minimize vulnerabilities - Providing only the hardware and software needed for optimal ECG workflow, MAC 5 reduces risks by eliminating pathways for breaches or attacks.
- Control access - Role-based LDAP user authentication, patient data encryption and secure network connections help ensure MAC 5 interacts only with trusted people and systems.
- Stay on top of emerging threats – The GE Healthcare cybersecurity team is constantly monitoring for threats and developing security enhancements for our customers.
General Specifications:
Full Specifications: MAC 5 Data Sheet & MAC 5 Lite Data Sheet
Instrument type:
- Microprocessor augmented automatic electrocardiograph; 10-lead wire acquisition with programmable lead configuration
ECG interpretation:
- Marquette™ 12SL™ ECG analysis program for adults and pediatrics
Computerized measurements:
- 12-lead analysis includes measurements
Heart rate meter:
- 30 to 300 BPM ±10% or 5 BPM, whichever is greater — heart rates outside this range will not be displayed
ECG data formats
External archiving:
Pre-acquisition:
- Provides 10 seconds of instantaneous ECG acquisition
Digital rhythm:
- Up to 5 minutes of continuous rhythm storage (exportable as a PDF)
Full disclosure:
- Review up to 5 minutes of 12 Lead ECG, abiliy to selet 10 second Resting ECG records, ability to generate 5 minute single lead full disclosure report
Storage:
- 300 records consisting of 10 second Resting ECG and rhythm records200 minutes Digital Rhythm or Full Disclosure records
Dynamic range:
- AC differential ± 10 mV DC offset ±600 mV
Common mode rejection:
- >125 dB (>100 dB with AC filter disabled)
Input impedance:
Defibrillation protection:
Patient leakage:
<10 μA
Specifications for digital acquisition and analysis of waveforms
Analog to digital conversion:
- 24-bit analog to digital conversion resolution
- Over sampled rate: 512 ksps
Down sampled ECG waveform:
- Bandwidth: 0.04 to 300 Hz*
- Sample rate: 2 ksps
- Resolution: 1.22 μV
Input to 12SL:
- Bandwidth: 0.04, 0.56 ZPD to 300
- Hz*Sample rate: 1 ksps
- Resolution: 4.88 μV
Additional report
- 20 Hz, 40 Hz, 100 Hz, 150 Hz or 300 Hz*filters
Specification for stored / transmitted waveforms
Digital rhythm waveform:
- Bandwidth: 0.04, 0.56 ZPD to 300 Hz*waveform
- Sample rate: 1000 sps
- Resolution: 4.88 μV
'12-lead ECG waveform:
- Bandwidth: 0.04, 0.56 ZPD to 300 Hz*
- Sample rate: 500 and 1000 sps
- Resolution: 4.88 μV
Representative (median) complex:
- Sample rate: 500 and 1000 sps
- Resolution: 4.88 μV
Pace detection
Pacemaker waveform
Pace detection
- Duration: 0.2 ms to 2.1 ms
- Amplitude: 2 mV to 700 mV
- Separation: 1 ms or greater
Pace annotation
- Dedicated pace channel on display and printed reports (configurable)