The Philips HeartStart Onsite Defibrillator is a light, well-designed product. When designing the HeartStart Onsite Defibrillator, Philips’ engineers were guided by exhaustive studies on user feedback, making the HeartStart Onsite defibrillator among the most user-friendly products on the market. Additional features of the Philips Heartstart Onsite defibrillator include:

  • 8 Year Mail in Warranty
  • lightweight defibrillator is easy to carry and transport 
  • With voice prompts, the HeartStart OnSite Defibrillator guides the caregiver through the process of defibrillation and CPR.  
  • SMART Pads pass data to the defibrillator, which is capable of recognizing and adapting to patient actions. The HeartStart Onsite AED Treats the most common cause of Sudden Cardiac Arrest by delivering a shock to the heart
  • HS1 battery includes a 4-year warranty from time of installation
  • 1 Pre-installed battery
  • 1 pre-installed SMART pads cartridge
Defibrillator:
  • Waveform: Truncated Exponential Biphasic.Waveform parameters adjusted as a function of each patient’s impedance.
  • Energy: Adult - Nominal 150 Joules into a 50 ohm load. Infant/Child - Nominal 50 Joules into a 50 ohm load. Automatically set based on type of SMART Pads cartridge installed.
  • Shock-to-Shock Cycle Time: Typically less than 20 seconds between shocks in a series
  • Quick Shock: Able to deliver a shock after the end of a CPR interval, typically in eight seconds
  • Voice Instructions: Detailed voice messages guide responder through use of the defibrillator
  • CPR Coaching: Instructions for adult and infant/child CPR available at user’s option
  • Shock Delivery: Via adhesive pads placed on patient’s bare skin as illustrated on pads
  • Controls: Green SMART Pads cartridge handle, green On/Off button, blue i-button,orange Shock button
  • Indicators: Ready light; blue i-button; caution light

  • Dimensions:
  • Depth: 7.4 in/19 cm
  • Width: 8.3 in/21 cm
  • Height: 2.8 in/7 cm
  • Weight: 3.3 lbs/1.5 kg, with battery and pads case

  • Environmental:
  • Sealing: Solid objects per EN60529 class IP2X; Drip-proof per EN60529 class IPX1
  • Temperature: Operating - 32º - 122º F (0º- 50º C); Standby - 50º - 109º F (10º- 43º C)
  • Humidity: Operating - 0% to 95% relative,non-condensing; Standby - 0% to 75% relative,non-condensing
  • Altitude: Operating - 0 to 15,000 feet; Standby - 0 to 8,500 feet > 48 hours and 8,500 to 15,000 feet < 48 hours
  • Shock/Drop Abuse: Withstands 1 meter drop to any edge, corner or surface
  • Vibration Meets EN1789 random and swept sine, road ambulance specification in operating and standby states
  • EMI (Radiated/Immunity) Meets EN55011 Group 1 Level B Class B and EN61000-4-3

  • Automated and User-Activated Self-Tests:
  • Daily Automatic Self-Tests: Tests internal circuitry, waveform delivery system, pads cartridge and battery capacity
  • Pads Integrity Test: Specifically tests readiness-for-use of pads (gel moisture)
  • Battery Insertion Test: Upon battery insertion, extensive automatic self-tests and user-interactive test check device readiness
  • Status Indicator: Blinking green “Ready” light indicates ready for use. Audible “chirp” indicates need for maintenance

  • Data Recording and Transmission:
  • Infrared: Wireless transmission of event data to a PC or Palm® PDA, using the IrDA protocol
  • Data Stored: First 15 minutes of ECG and the entire incident’s events and analysis decisions